Medical MEB-1K5A24T AAA Product Change Notice


Delta reduces leakage current to increase user safety for MEB-1K5A24T AAA series medical power supplies. Please refer to the table below for information on the changes. Comprehensive testing and verification of the results have been conducted to ensure that all customers can expect consistent, high-quality product performance after these design changes.


If you require any further clarification, please refer to the latest specifications on the DeltaPSU website or reach out to us via our Contact Us Form.


Model List:


Model Name Description Old Revision
New Revision
Implementation Date
MEB-1K5A24T AAA Reduce leakage current 00 01 2025/10/16

Modification of MEB-1K5A24T AAA:


Specifications Before the Change Specifications After the Change
Touch Current (maximum) 0.1 mA @ Normal Condition
0.5 mA @ Single Fault Condition
0.1 mA @ Normal Condition
0.2 mA @ Single Fault Condition
Earth Leakage Current (maximum) 0.5 mA @ Normal Condition
1.0 mA @ Single Fault Condition
0.2 mA @ Normal Condition
0.4 mA @ Single Fault Condition

İlgili Haberler

PJ Series Product Upgrade Notice (PCN-23-005)

Due to the impact of global material and component shortages, Delta PJ series power suppli ...

Daha Fazla

Product Serial Number Format Change Notice (PCN-23-001)

Delta product serial number format will add one more digit (18 digits → 19 digits) to iden ...

Daha Fazla

DRP024V240W1AA Product Upgrade Notice

Due to end-of-life of certain key components, DRP024V240W1AA model will revise the out ...